{"_buckets": {"deposit": "3565f88f-9a31-4877-a62e-df974771a36d"}, "_deposit": {"created_by": 11, "id": "106", "owners": [11], "pid": {"revision_id": 0, "type": "depid", "value": "106"}, "status": "published"}, "_oai": {"id": "oai:yachts.repo.nii.ac.jp:00000106", "sets": ["47"]}, "author_link": ["2133", "2132", "2135", "2134", "2131", "2136"], "item_2_biblio_info_6": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2002-03", "bibliographicIssueDateType": "Issued"}, "bibliographicPageEnd": "35", "bibliographicPageStart": "31", "bibliographicVolumeNumber": "5", "bibliographic_titles": [{"bibliographic_title": "山形保健医療研究"}]}]}, "item_2_description_4": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "山形県立保健医療大学及びUniversity of Colorado Health Science Center rehabilitation Medicine,Physical Therapy Program所有の三次元動作解析装置によって得られるデータの精度を検討した.測定は両大学において正三角形のアルミフレームを回転台に固定し,静止時と回転時にアルミフレームの3個所に貼り付けた反射マーカーによって成す角度を計測することで行った.その結果,両大学とも十分に信頼性のある測定が可能であったことから,データに互換性があることが示された.今後,両大学における共同研究が期待できるものと思われた", "subitem_description_type": "Abstract"}]}, "item_2_relation_29": {"attribute_name": "医中誌ID", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "2002215062", "subitem_relation_type_select": "ICHUSHI"}}]}, "item_2_source_id_7": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "1343-876X", "subitem_source_identifier_type": "ISSN"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "小野, 武也"}], "nameIdentifiers": [{"nameIdentifier": "2131", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "吉野, 直美"}], "nameIdentifiers": [{"nameIdentifier": "2132", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "宮崎, 純弥"}], "nameIdentifiers": [{"nameIdentifier": "2133", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "鈴木, 克彦"}], "nameIdentifiers": [{"nameIdentifier": "2134", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "駒沢, 治夫"}], "nameIdentifiers": [{"nameIdentifier": "2135", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "伊橋, 光二"}], "nameIdentifiers": [{"nameIdentifier": "2136", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2016-11-07"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "2002.05.pdf", "filesize": [{"value": "1.6 MB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 1600000.0, "url": {"label": "2002.05", "url": "https://yachts.repo.nii.ac.jp/record/106/files/2002.05.pdf"}, "version_id": "036893cd-d4c0-4583-961c-52bc1f99c94b"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "運動障害(診断)", "subitem_subject_scheme": "Other"}, {"subitem_subject": "機器と資材用品", "subitem_subject_scheme": "Other"}, {"subitem_subject": "大学", "subitem_subject_scheme": "Other"}, {"subitem_subject": "動作", "subitem_subject_scheme": "Other"}, {"subitem_subject": "理学療法", "subitem_subject_scheme": "Other"}, {"subitem_subject": "三次元", "subitem_subject_scheme": "Other"}, {"subitem_subject": "測定機器", "subitem_subject_scheme": "Other"}, {"subitem_subject": "動作分析", "subitem_subject_scheme": "Other"}, {"subitem_subject": "測定精度", "subitem_subject_scheme": "Other"}, {"subitem_subject": "ヒト", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "jpn"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "departmental bulletin paper", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "コロラド大学と当大学における三次元動作解析装置精度の検討", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "コロラド大学と当大学における三次元動作解析装置精度の検討"}, {"subitem_title": "The reliability of the three-dimensional motion analysis system in the University of Colorado Health Science Center,Rehabilitation Medicine,Physical Therapy Program and Yamagata Prefectural University of Health Science.", "subitem_title_language": "en"}]}, "item_type_id": "2", "owner": "11", "path": ["47"], "permalink_uri": "https://yachts.repo.nii.ac.jp/records/106", "pubdate": {"attribute_name": "公開日", "attribute_value": "2016-11-07"}, "publish_date": "2016-11-07", "publish_status": "0", "recid": "106", "relation": {}, "relation_version_is_last": true, "title": ["コロラド大学と当大学における三次元動作解析装置精度の検討"], "weko_shared_id": -1}
コロラド大学と当大学における三次元動作解析装置精度の検討
https://yachts.repo.nii.ac.jp/records/106
https://yachts.repo.nii.ac.jp/records/1064a4fa078-4562-4c0e-a853-9876b0eb684b
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2016-11-07 | |||||
タイトル | ||||||
タイトル | コロラド大学と当大学における三次元動作解析装置精度の検討 | |||||
タイトル | ||||||
言語 | en | |||||
タイトル | The reliability of the three-dimensional motion analysis system in the University of Colorado Health Science Center,Rehabilitation Medicine,Physical Therapy Program and Yamagata Prefectural University of Health Science. | |||||
言語 | ||||||
言語 | jpn | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 運動障害(診断) | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 機器と資材用品 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 大学 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 動作 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 理学療法 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 三次元 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 測定機器 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 動作分析 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 測定精度 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | ヒト | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
著者 |
小野, 武也
× 小野, 武也× 吉野, 直美× 宮崎, 純弥× 鈴木, 克彦× 駒沢, 治夫× 伊橋, 光二 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | 山形県立保健医療大学及びUniversity of Colorado Health Science Center rehabilitation Medicine,Physical Therapy Program所有の三次元動作解析装置によって得られるデータの精度を検討した.測定は両大学において正三角形のアルミフレームを回転台に固定し,静止時と回転時にアルミフレームの3個所に貼り付けた反射マーカーによって成す角度を計測することで行った.その結果,両大学とも十分に信頼性のある測定が可能であったことから,データに互換性があることが示された.今後,両大学における共同研究が期待できるものと思われた | |||||
書誌情報 |
山形保健医療研究 巻 5, p. 31-35, 発行日 2002-03 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1343-876X | |||||
医中誌ID | ||||||
識別子タイプ | ICHUSHI | |||||
関連識別子 | 2002215062 |